60W open frame AC-DC medical-grade power supply ACMS61B
PRODUCT PICTURE
DESCRIPTION
The purpose of the document is to specify the functional requirements of a ACMS61B-120500 medical switching power
supply.
/MODEL NAME RULE
ACMS61B-aaabbb
-ACMS=Single-channel medical switching power supply
-61=Name of the series
-aaa= "120"stands for the output voltage."120" means 12.0V.
-bbb="500"stands for the output voltage."500" means 5.0A.
AC Leakage Current
Normal touch current less than 0.1mA at 264Vac input.
Single fault touch current less than 0.5mA at 264Vac input.
Normal earth leakage current less than 5mA at 264Vac input.
Single fault earth leakage current less than 10mA at 264Vac input.
PROTECTION FUNCTION
6.1 Short Circuit Test
The power supply will be auto recovering when the short circuit faults removed.
6.2 Over Current Protection
When the output current reaches 110%~200% of the rated current, the power supply will be protected. After
the over current fault is removed, the power supply will be automatically recovering.
SAFETY REQUIREMENT
Safety: accord with.
The power supply is designed for indoor use to meet IEC 60601-1/UL60601-1and EN 60601-1standards and for
the following safety regulations.
Dielectric Strength Hi-Pot
The primary to secondary: 4000Vac 5mA for 60 seconds
The primary to earth: 1500Vac 5mA for 60 seconds
The secondary to earth:500Vdc 5mA for 60 seconds
parameter
Feature |
Model |
Parameters(MultipleOutput) |
|
ACMM46 |
Output Voltage |
+12V |
|
|
|
||
Output Current |
3A |
||
|
|
||
Output Voltage |
+5V |
||
|
|
||
Output Current |
2A |
Application
The 46W Dual Output Medical Power Supply plays a crucial role in powering a wide range of medical equipment and devices, supporting various clinical applications and contributing to the delivery of high-quality healthcare services.