40W destop medical adapter UL CE Switching Power Supply Adaptor 40W 5V 6V 9V 12V 15V 19V 24V LXCP40
Input Characteristics
Input current
The maximum input current is 1.0a at nominal ac input voltage and full load.
Inrush current
The inrush current will not exceed 50a at 264vac input for a cold start at 25℃.
Static loss
The static loss is less than 0.5w at normal input voltage.
Efficiency
The efficiency shall be higher than 80% while measuring at 115vac&230vac input voltage, full load.
Output Characteristics
Item |
Model number |
Output characteristics |
|||||
|
LXCP40 series |
LXCP40(II) series |
Rated output voltage(V) |
Output voltage range(V) |
Rated output current(A) |
Rated output power(W) |
|
Model number |
LXCP40-005 |
LXCP40-005G |
LXCP40(II)-005 |
5.0 |
4.60~5.40 |
4.00 |
20.0 |
LXCP40-009 |
LXCP40-009G |
LXCP40(II)-009 |
9.0 |
8.60~9.40 |
4.00 |
36.0 |
|
LXCP40-012 |
LXCP40-012G |
LXCP40(II)-012 |
12.0 |
11.40~12.60 |
3.33 |
40.0 |
|
LXCP40-015 |
LXCP40-015G |
LXCP40(II)-015 |
15.0 |
14.20~15.80 |
2.67 |
40.0 |
|
LXCP40-168 |
LXCP40-168G |
LXCP40(II)-168 |
16.8 |
15.95~17.64 |
2.38 |
40.0 |
|
LXCP40-018 |
LXCP40-018G |
LXCP40(II)-018 |
18.0 |
17.10~18.90 |
2.22 |
40.0 |
|
LXCP40-019 |
LXCP40-019G |
LXCP40(II)-019 |
19.0 |
18.05~19.95 |
2.10 |
40.0 |
|
LXCP40-020 |
LXCP40-020G |
LXCP40(II)-020 |
20.0 |
19.00~21.00 |
2.00 |
40.0 |
|
LXCP40-024 |
LXCP40-024G |
LXCP40(II)-024 |
24.0 |
23.28~24.72 |
1.67 |
40.0 |
|
LXCP40-026 |
LXCP40-026G |
LXCP40(II)-026 |
26.0 |
25.22~26.78 |
1.55 |
40.0 |
|
LXCP40-028 |
LXCP40-028G |
LXCP40(II)-028 |
28.0 |
27.16~28.84 |
1.42 |
40.0 |
|
LXCP40-030 |
LXCP40-030G |
LXCP40(II)-030 |
30.0 |
29.10~30.90 |
1.33 |
40.0 |
|
LXCP40-032 |
LXCP40-032G |
LXCP40(II)-032 |
32.0 |
31.04~32.96 |
1.25 |
40.0 |
|
LXCP40-036 |
LXCP40-036G |
LXCP40(II)-036 |
36.0+ |
34.92~37.08 |
1.11 |
40.0 |
|
LXCP40-040 |
LXCP40-040G |
LXCP40(II)-040 |
40.0 |
38.80~41.20 |
1.00 |
40.0 |
|
LXCP40-042 |
LXCP40-042G |
LXCP40(II)-042 |
42.0 |
40.74~43.26 |
0.95 |
40.0 |
|
LXCP40-046 |
LXCP40-046G |
LXCP40(II)-046 |
46.0 |
44.62~47.38 |
0.87 |
40.0 |
|
LXCP40-048 |
LXCP40-048G |
LXCP40(II)-048 |
48.0 |
46.56~49.44 |
0.84 |
40.3 |
|
Voltage regulation |
±2% |
||||||
Overshoot |
≤5% |
||||||
Hold up time |
≥10 ms |
||||||
Efficiency |
≥80%(≥68%@5V) |
||||||
Over current protection(A) |
105-200%Vout |
||||||
Over voltage protection(V) |
105-125%Iout |
||||||
Ripple and noise |
≤100mV (measured by 20MHz bandwidth oscilloscope and terminated each output with a 0.1uF ceramic capacitor and a10uF aluminum electrolytic capacitor)(>24V:≤1% Vo) |
||||||
Over temperature protection |
The power supply shall go into stop mode when the IC's temperature is over its trigger point and when IC temperature is lower than regulation value the power supply shall be self-recovering. |
||||||
Short circuit protection |
The power supply shall withstand continuous current of short circuit on output no any damage. When the fault condition is removed the power supply shall be self-recovering. |
The power supply shall be compliant with each item in this specification for the following environmental conditions.
Ambient temperature |
-10℃~40℃ |
Relative humidity |
20%~90% |
Storage Condition
Ambient temperature |
-20℃~80℃ |
Relative humidity |
10%~90% |
MTBF
The power supply shall have a minimum predicted MTBF(MIL-STD-217F) of 100000 hours at normal operation condition and normal use.
WEIGHT
The weight of the power supply shall be about 200g.
Electromagnetic Compatiblity
Numbe |
Item |
Specifications |
Class |
Standards |
1 |
(CE) |
CLASS B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18 |
2 |
(RE) |
CLASS B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18 |
3 |
(SURGE) |
Line to line±1KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-5; GB17626.5 |
|
|
Line to GND±2KV |
A |
|
4 |
(ESD) |
Air discharge±15KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-2; GB17626.2 |
|
|
Contact discharge±8KV |
A |
|
5 |
(EFT/B) |
±2KV (BURST FREQUENCY=100KHZ) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-4; GB17626.4 |
6 |
(DIP) |
Dropped to 0%Ut,last 5000ms(250cycle) |
B |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-11; GB17626.11 |
Dropped to 30%Ut,last 500ms(25 cycle) |
B |
|
||
Dropped to 0%Ut,last 20ms(1 cycle) |
B |
|
||
Dropped to0%Ut,last 10ms(0.5 cycle) |
A |
|
||
7 |
(RS) |
Test frequency:80MHz~2700MHz; Field intensity:10V/m;80%AM(1KHz) Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-3; GB17626.3 |
8 |
(CS) |
Test frequency:0.15MHz~80MHz; Field intensity:6Vrms; Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-6; GB17626.6 |
9 |
(THD) |
CLASSA(in system) |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-2; GB17625.1 |
10 |
Voltage Fluctuation and Flicker |
Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4% |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-3; GB17625.2 |
11 |
Power Frequency Magnetic Field |
30A/m |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-8; GB17626.8 |
Safety: accord with
Item |
Country |
Standard |
|
□ |
UL |
USA |
UL60601-1 |
□ |
CE |
Europe |
EN60601-1 |
□ |
CB |
Worldwide |
IEC60601-1 |
□ |
TUV |
Germany |
IEC60601-1 |
□ |
NRTL |
CUSA |
IEC60601-1/UL60601-1 |
□ |
GS |
Germany |
EN60601-1 |
□ |
BS |
England |
EN60601-1 |
□ |
SAA |
Australia |
AS/NZS6-1 |
Mechanical Requirement
OUTLINE AND LABEL ARTWORK
Note: The figure only lists part of the optional DC plug diagram, the details are selected from the LXC_ cable statistics workbook xlsx file.