Normative reference
IEC68-2-1 IEC68-2-1A GB2423.1
Basic environmental testing procedures for electric and electronic products - Test A: Low temperature test method;
IEC68-2-2 IEC68-2-2A GB2423.2
Basic environmental testing procedures for electric and electronic products - Test B: Dry hear;
IEC68-2-3 GB2423.9
Basic environmental test procedure for electrical and electronic products Test Cd: Damp heat, steady state, primarily for equipment
IEC68-2-6 GB2423.10
Basic environmental testing procedures for electric and electronic products Test Fc and guidance:Vibration sinusoidal
IEC60601-1 GB 9706.1
Medical electrical equipment - part 1: general requirements for safety;
IEC60601-1-2 YY0505
IEC60601-1-2 YY0505Medical electrical equipment - part 1-2: common safety requirements and standards: electromagnetic compatibility requirements and tests.
Input characteristics
Number |
Item |
Specifications |
Unit |
Note |
1 |
Normal input |
220 |
Vac |
|
2 |
Range |
100~240 |
Vac |
|
3 |
Frequencry |
50/60 |
Hz |
|
4 |
Current max |
0.7 |
A |
90Vac,25℃ |
5 |
Inrush current |
≤60 |
A |
220Vac,25℃ |
≤30 |
|
115Vac,25℃ |
Item |
Model number |
OUTPUT CHARACTERISTICS |
||
Rated output voltage (v) |
Output voltage range (V) |
Output currentrange (A) |
||
Model number(High power version with radiator) |
ACMS23-005 |
5.0 |
4.70~5.30 |
0-4.00 |
ACMS23-075 |
7.5 |
7.20~7.80 |
0-3.00 |
|
ACMS23-009 |
9.0 |
8.55~9.45 |
0-3.00 |
|
ACMS23-012 |
12.0 |
11.64~12.36 |
0-2.50 |
|
ACMS23-136 |
13.6 |
13.30~13.90 |
0-2.20 |
|
ACMS23-142 |
14.2 |
13.90~14.50 |
0-2.10 |
|
ACMS23-015 |
15.0 |
14.70~15.30 |
0-2.00 |
|
ACMS23-168 |
16.8 |
16.46~17.13 |
0-1.78 |
|
ACMS23-018 |
18.0 |
17.80~18.20 |
0-1.66 |
|
ACMS23-019 |
19.0 |
18.70~19.30 |
0-1.57 |
|
ACMS23-020 |
20.0 |
19.70~20.30 |
0-1.50 |
|
ACMS23-024 |
24.0 |
23.50~24.50 |
0-1.30 |
|
Model number(version without radiator)
|
ACMS23-005 |
5.0 |
4.70~5.30 |
0-3.00 |
ACMS23-075 |
7.5 |
7.20~7.80 |
0-2.50 |
|
ACMS23-009 |
9.0 |
8.55~9.45 |
0-2.50 |
|
ACMS23-012 |
12.0 |
11.64~12.36 |
0-2.10 |
|
ACMS23-136 |
13.6 |
13.30~13.90 |
0-1.83 |
|
ACMS23-142 |
14.2 |
13.90~14.50 |
0-1.76 |
|
ACMS23-015 |
15.0 |
14.70~15.30 |
0-1.67 |
|
ACMS23-168 |
16.8 |
16.46~17.13 |
0-1.49 |
|
ACMS23-018 |
18.0 |
17.80~18.20 |
0-1.39 |
|
ACMS23-019 |
19.0 |
18.70~19.30 |
0-1.32 |
|
ACMS23-020 |
20.0 |
19.70~20.30 |
0-1.25 |
|
ACMS23-024 |
24.0 |
23.50~24.50 |
0-1.10 |
|
load regulation |
±3% |
|||
voltage regulation |
±3% |
|||
overshoot |
≤5% |
|||
Hold up time |
≥10 mS |
|||
Ripple and noise |
≤100mV (measured by 20MHz bandwidth oscilloscope and terminated each output with a 0.1uF ceramic capacitor and a10uF aluminum electrolytic capacitor) |
|||
Over voltage protection |
110-150%Vout |
|||
Over current protection |
110-170%Iout |
|||
Over temperature protection |
The power supply shall go into stop mode when the IC's temperature is over its trigger point and when IC temperature is lower than regulation value, the power supply shall be self-recovering. |
|||
Short circuit protection |
The power supply shall withstand continuous current of short circuit on output no any damage, and when the fault condition is removed, the power supply shall be self-recovering. |
Electromagnetic Compatiblity
Numbe |
Item |
Specifications |
Class |
Standards |
1 |
(CE) |
CLASS B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18 |
2 |
(RE) |
CLASS B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18 |
3 |
(SURGE) |
Line to line±1KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-5; GB17626.5 |
|
|
Line to GND±2KV |
A |
|
4 |
(ESD) |
Air discharge±15KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-2; GB17626.2 |
|
|
Contact discharge±8KV |
A |
|
5 |
(EFT/B) |
±2KV (BURST FREQUENCY=100KHZ) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-4; GB17626.4 |
6 |
(DIP) |
Dropped to 0%Ut,last 5000ms(250cycle) |
B |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-11; GB17626.11 |
Dropped to 30%Ut,last 500ms(25 cycle) |
B |
|
||
Dropped to 0%Ut,last 20ms(1 cycle) |
B |
|
||
Dropped to0%Ut,last 10ms(0.5 cycle) |
A |
|
||
7 |
(RS) |
Test frequency:80MHz~2700MHz; Field intensity:10V/m;80%AM(1KHz) Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-3; GB17626.3 |
8 |
(CS) |
Test frequency:0.15MHz~80MHz; Field intensity:6Vrms; Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-6; GB17626.6 |
9 |
(THD) |
CLASSA(in system) |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-2; GB17625.1 |
10 |
Voltage Fluctuation and Flicker |
Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%; Maximum relative voltage variation (dmax) under4% |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-3; GB17625.2 |
11 |
Power Frequency Magnetic Field |
30A/m |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-8; GB17626.8 |
Outline