20W medical power supply module 9V 12V 15V 24V open frame reliable switching power supply ACMS25E
Input Characteristics
IEC68-2-1 IEC68-2-1A GB2423.1-1989
Basic environmental testing procedures for electric and electronic products - Test A: Low temperature test method;
IEC68-2-2 IEC68-2-2A GB2423.2-1989
Basic environmental testing procedures for electric and electronic products - Test B: Dry hear;
IEC68-2-3 GB2423.9-1989
Basic environmental test procedure for electrical and electronic products Test Cd: Damp heat, steady state, primarily for equipment
IEC68-2-6 GB2423.10-1995
Basic environmental testing procedures for electric and electronic products Test Fc and guidance:Vibration sinusoidal
IEC60601-1 GB 9706.1—2007
Medical electrical equipment - part 1: general requirements for safety;
IEC60601-1-2 YY0505-2012
Medical electrical equipment - part 1-2: common safety requirements and standards: electromagnetic compatibility requirements and tests.
Ambient condition
Environment of work condition
Number |
Item |
Specifications |
Unit |
Note |
1 |
Work temp |
0~+40 |
℃ |
|
2 |
Relative humidity |
10~90 |
% |
No condensation |
3 |
Atmospheric pressure |
70K~106K |
Pa |
Normal |
4 |
Height |
≤5000 |
m |
Work temp descend 1℃ every 300m over 5000m |
5 |
Cooling Mode |
Free air convection |
|
Environment of storage condition
Number |
Item |
Specifications |
Unit |
Note |
1 |
Storage temp |
-20~+80 |
℃ |
No work |
2 |
Storage humidity |
10~90 |
% |
No condensation |
Environment of test condition
Number |
Item |
Specifications |
Note |
1 |
High temp |
+40℃ |
Work normal and performance good |
2 |
Low temp |
0℃ |
Work normal and performance good |
3 |
Storage in High temp |
+80℃ |
Work normal and performance good after 2 hours in normal Environment |
4 |
Storage in low temp |
-20℃ |
Work normal and performance good after 2 hours in normal Environment |
5 |
Vibration |
The sample can afford for 5~500HZ free vibration in X Y Z three direction. spectral density is 10 m2/s3 in range of 5~10Hz,3m2/s3 in range of 10~200HZ,1 m2/s3 in range of 200~500HZ.20min every direction. |
normal |
6 |
Shock |
Half sine wave,peak accelerated speed 20G,last 11mS,6faces,3times |
normal |
Electromagnetic Compatiblity
Numbe |
Item |
Specifications |
Class |
Standards |
1 |
(CE) |
CLASS B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18 |
2 |
(RE) |
CLASS B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18 |
3 |
(SURGE) |
Line to line±1KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-5; GB17626.5 |
|
|
Line to GND±2KV |
A |
|
4 |
(ESD) |
Air discharge±15KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-2; GB17626.2 |
|
|
Contact discharge±8KV |
A |
|
5 |
(EFT/B) |
±2KV (BURST FREQUENCY=100KHZ) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-4; GB17626.4 |
6 |
(DIP) |
Dropped to 0%Ut,last 5000ms(250cycle) |
B |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-11; GB17626.11 |
Dropped to 30%Ut,last 500ms(25 cycle) |
B |
|
||
Dropped to 0%Ut,last 20ms(1 cycle) |
B |
|
||
Dropped to0%Ut,last 10ms(0.5 cycle) |
A |
|
||
7 |
(RS) |
Test frequency:80MHz~2700MHz; Field intensity:10V/m;80%AM(1KHz) Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-3; GB17626.3 |
8 |
(CS) |
Test frequency:0.15MHz~80MHz; Field intensity:6Vrms; Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-6; GB17626.6 |
9 |
(THD) |
CLASSA(in system) |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-2; GB17625.1 |
10 |
Voltage Fluctuation and Flicker |
Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4% |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-3; GB17625.2 |
11 |
Power Frequency Magnetic Field |
30A/m |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-8; GB17626.8 |
Safety: accord with
Number |
Item |
Specifications |
1 |
Fault impact |
The power supply will left in fault and no harmful to anthing else. |
2 |
Components |
The components in the device above industrial leve |
3 |
Audible noise |
45dB(A),distance 1m |
4 |
Cooling Mode |
Free air convection |