180W Open Frame Six Output Medical Power Supply 24V 12V5V 12V -12V +110V ACMM180
PRODUCT PICTURE
1.1 ACMM180
1.2 ACMM180A
1.3 ACMM180B
1.4 ACMM180C
1.5 ACMM180D
SERIES TABLE
ACMM180X Typical Parameter Table | |||||||
Voltage |
Output |
||||||
|
V1 |
V2 |
V3 |
V4 |
V5 |
V6 |
ACMS43- 120 |
|
+24V |
+12V |
+5V |
+12V |
-12V |
+110V |
+12V |
Current |
0-4A |
0.3-3A |
0.5-3A |
0-1A |
0-1A |
0-0.1A |
0-4.2A |
ACMM180 |
√ |
√ |
√ |
√ |
√ |
√ |
|
ACMM180A |
√ |
√ |
√ |
√ |
√ |
√ |
|
ACMM180B |
√ |
√ |
√ |
√ |
√ |
√ |
√ |
ACMM180C |
√ |
√ |
√ |
√ |
√ |
√ |
√ |
ACMM180D |
√ |
√ |
√ |
√ |
√ |
√ |
PRODUCT OVERVIEW
This power supply is a multi-channel output AC/DC power supply, the whole power supply adopts the latest circuit design, with compact structure, reliable performance, the whole machine high precision voltage regulation characteristics. Output with short circuit protection, over current protection, over voltage protection and other functions.
It has good electromagnetic compatibility and meets the safety requirements of medical equipment.
STANDARDS REFERENCE
GB2423.1 Basic environmental test procedures for electrical and electronic products -- Test Ad: low temperature test method;
GB2423.2 Basic environmental test procedures for electrical and electronic products -- Test Bd: high temperature test method;
GB2423.9 Basic environmental test procedures for electrical and electronic products -- Test Cb: constant damp heat test method;
GB2423.10 Basic environmental test procedures for electrical and electronic products -- Test Fc: vibration test method;
GB 9706.1 (IDT IEC60601-1) medical electrical equipment - Part 1: Safety general requirements;
YY 0505(IDT IEC60601-1-2) medical electrical equipment - Part 1-2: Safety common requirements juxtaposed standards: electromagnetic compatibility requirements and tests.
Electromagnetic Compatiblity
Numbe |
Item |
Specifications |
Class |
Standards |
1 |
(CE) |
CLASS B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18 |
2 |
(RE) |
CLASS B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18 |
3 |
(SURGE) |
Line to line±1KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-5; GB17626.5 |
|
|
Line to GND±2KV |
A |
|
4 |
(ESD) |
Air discharge±15KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-2; GB17626.2 |
|
|
Contact discharge±8KV |
A |
|
5 |
(EFT/B) |
±2KV (BURST FREQUENCY=100KHZ) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-4; GB17626.4 |
6 |
(DIP) |
Dropped to 0%Ut,last 5000ms(250cycle) |
B |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-11; GB17626.11 |
Dropped to 30%Ut,last 500ms(25 cycle) |
B |
|
||
Dropped to 0%Ut,last 20ms(1 cycle) |
B |
|
||
Dropped to0%Ut,last 10ms(0.5 cycle) |
A |
|
||
7 |
(RS) |
Test frequency:80MHz~2700MHz; Field intensity:10V/m;80%AM(1KHz) Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-3; GB17626.3 |
8 |
(CS) |
Test frequency:0.15MHz~80MHz; Field intensity:6Vrms; Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-6; GB17626.6 |
9 |
(THD) |
CLASSA(in system) |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-2; GB17625.1 |
10 |
Voltage Fluctuation and Flicker |
Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4% |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-3; GB17625.2 |
11 |
Power Frequency Magnetic Field |
30A/m |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-8; GB17626.8 |
OUTLINE
Number |
Item |
Specifications |
Note |
1 |
The failure to isolate |
After the power module fails, there is no output. The power module is reliably isolated and does not cause any damage to the device or personnel |
If a power module fails, the mains power supply system cannot be affected |
2 |
Acoustic noise |
≦45dB(A), test distance: 1m |
Mechanical Requirement
Label
The packing case is marked with the product name, model, manufacturer's logo, manufacturer's quality department's inspection certificate, manufacturing date, etc.
The product specification and attached list are attached in the packing case.