15W Factory OEM TUV CE EMC UK CA AC adaptor 12v 1a power supply adapter for Nebulizers LXCP12X
Input Characteristics
Input Voltage:
Nominal Voltage: 100~240Vac
Variation Range: 90~264Vac
Input Frequency:
Nominal Frequency: 50/60Hz.
Variation Frequency: 47~63Hz
Input Current:
0.5Amps Max. at any input voltage and rated DC output and rated load.
Inrush Current:
50Amps max. cold start at 264Vac input, with rated load and 25℃ ambient.
AC Leakage Current:
Normal 0.1mA Max.at 264Vac input.
Single fault 0.2mA Max.at 264Vac input.
Output Characteristics
Model name |
Output voltage(V) |
Output range(V) |
Output current (A) |
Output power (W) |
LXCP12X-036 |
3.6 |
3.10~4.00 |
2.25 |
8.10 |
LXCP12X-042 |
4.2 |
3.70~4.60 |
2.25 |
9.45 |
LXCP12X-050 |
5.0 |
4.40~5.60 |
2.25 |
11.25 |
LXCP12X-060 |
6.0 |
5.40~6.60 |
2.25 |
13.50 |
LXCP12X-075 |
7.5 |
7.00~8.10 |
1.60 |
12.00 |
LXCP12X-090 |
9.0 |
8.50~9.70 |
1.50 |
13.50 |
LXCP12X-100 |
10.0 |
9.50~10.70 |
1.20 |
12.00 |
LXCP12X-110 |
11.0 |
10.30~11.70 |
1.00 |
11.00 |
LXCP12X-120 |
12.0 |
11.30~12.70 |
1.00 |
12.00 |
LXCP12X-135 |
13.5 |
12.80~14.20 |
1.00 |
13.50 |
LXCP12X-136 |
13.6 |
12.90~14.30 |
1.00 |
13.60 |
LXCP12X-150 |
15.0 |
14.30~15.70 |
1.00 |
15.00 |
LXCP12X-168 |
16.8 |
16.10~17.50 |
0.90 |
15.12 |
Line regulation |
±5% |
|||
Load regulation |
±8% |
Ripple and Noise
Test condition:Under nominal voltage and rated load, the ripple and noise are as follows when measure with max. bandwidth of 20MHz and Parallel 10uF/0.1uF, crossed connected at testing point. Ripple and noise(max.) less than 200mVp-p.
Turn On Delay Time
3 second max.at 220Vac input and output rated load.
Rise Time
50ms max.at 115Vac input and output rated load.
Hold Up Time:
10ms min.at 115Vac input and output rated Load.
Efficiency:
70% Min. at 115/230Vac input voltage, rated load.
Protection Function
Short circuit test
The power supply will be auto recovering when short circuit faults removed.
Over current Protection
When the output current reaches 110%~200% of the rated current, the power supply will be protected. After the over current fault is removed, the power supply will be automatically recovering.
Over Voltage Protection
The power supply shall be stop when output voltage reaches its over-voltage condition trigger point. When the fault condition is removed the power supply shall be self-recovering. The over voltage protection function is built-in the PWM IC of the power supply.
Over TEMP Protection
The power supply shall go into stop mode when the IC's temperature is over its trigger point and when IC temperature is lower than regulation value, the power supply shall be self-recovering.
Environmental Requiremet
Operating Temperature
0℃ to 40℃,Rated load, Normal operation.
Storage Temperature:-20℃ to 80℃
No work
Storage Humidity
10%~90%RH, No condensation.
Atmospheric pressure
54-106KPa, Normal
Altitude
5000M,work temp descend 1℃ every 500m over 5000m.
(9~200Hz,Acceleration 5m/S2)
Transportation: 5-9Hz, A=3.5mm
Acceleration=5m/S2
Acceleration=15m/S2
Axes, 10 cycles per axis
No permanent damage may occur during testing.
The Power supply can be restore to normal after power off/on.
Dropping Packed
1m for wall mount type and 760mm for desktop type as above described.
The horizontal surface consists of hardwood at least 13mm thick, mounted on two layers of plywood and away from land 19mm ~ 20mm.
Axial Vibration
Axes,10 cycles per axis. No permanent damage may occur during testing. The power supply can be recovery to normal after power off/o
Drop Test
1M for wall mounting type and 760mm for desktop type as above described. The horizontal surface consists of the hardwood at least 13mm thick, mounted on two layers of plywood and away from land 19mm ~ 20mm.
MTBF
The power supply shall have a minimum predicted MTBF(MIL-STD-217F) of 100000 hours at normal operation condition and normal use.
Electromagnetic Compatiblity
Number |
Item |
Specifications |
Class |
Standards |
1 |
CE |
Class B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18; GB4824 |
2 |
RE |
Class B |
/ |
IEC/EN60601-1-2;YY0505 GB4824; EN55011; FCC Part 18 |
3 |
SURGE |
Line to line ± 1KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-5; GB17626.5 |
|
|
Line to GND ± 2KV |
/ |
|
4 |
ESD |
Air discharge ± 15KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-2; GB17626.2 |
|
|
Contact discharge ± 8KV |
A |
|
5 |
EFT/B |
± 2KV (Burst frequency=100KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-4; GB17626.4 |
6 |
DIP |
Dropped to 0%Ut,last 5000ms(250cycle) |
B |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-11 GB17626.11 |
Dropped to 70%Ut,last 500ms(25 cycle) |
A |
|
||
Dropped to 0%Ut,last 20ms(1 cycle) |
A |
|
||
Dropped to0%Ut,last 10ms(0.5 cycle) |
A |
|
||
7 |
RS |
Test frequency:80MHz~2700MHz; Field intensity:10V/m; Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-3 GB17626.3 |
8 |
CS |
Test frequency:0.15MHz~80MHz; Field intensity:6Vrms; Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-6 GB17626.6 |
9 |
THD |
Class A(in system) |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-2; GB17625.1 |
10 |
Voltage Fluctuation and Flicker |
Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (Dmax) under4% |
/ |
IEC/EN60601-1-2 YY0505 IEC/EN61000-3-3 GB17625.2 |
11 |
Power Frequency Magnetic Field |
30A/m |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-8; GB17626.8 |
Safety: accord with
Item |
Country |
Standard |
|
□ |
UL |
USA |
UL60950-1/UL60601-1 |
R |
CE |
Europe |
EN60950-1/EN60601-1 |
R |
CB |
Worldwide |
IEC60601-1 |
□ |
TUV |
Germany |
IEC60601-1 |
R |
NRTL |
CUSA |
IEC60601-1/UL60601-1 |
□ |
GS |
Germany |
EN60601-1 |
□ |
BS |
England |
EN60601-1 |
□ |
SAA |
Australia |
AS/NZS6-1 |
Mechanical Requirement
![JF8%B)%]PN$9@)19P`GP@VFu09](https://ecdn6.globalso.com/upload/p/1150/image_other/2024-04/660b73122aeb088180.png)
Cable




label size:24.5mm*17.5mm
![LOA}S]7MQMXF)QSH$A(Q{E8vo2](https://ecdn6.globalso.com/upload/p/1150/image_other/2024-04/660b75453678746846.jpg)