12W detachable plug US EU AU UK adapter 5V 9V 12V wall mount adapter medical device adapter LXCP12
Input Characteristics
Input Voltage:
Nominal Voltage: 100~240Vac
Variation Range: 90~264Vac
Input Frequency:
Nominal Frequency: 50/60Hz.
Variation Frequency: 47~63Hz
Input Current:
0.5Amps Max. at any input voltage and rated DC output and rated load.
Inrush Current:
60Amps Max. cold start at 264Vac input, with rated load and 25℃ ambient.
AC Leakage Current:
Normal status,the Max. contacted current is less than 0.1mA when input voltage is 264Vac.
Single fault status,the Max. contacted current is less than 0.5mA when input voltage is 264Vac.
Output Characteristics
Model name |
Output Voltage(V) |
Rated Load (A) |
Output Range(V) |
Over current protection(A) |
Output power(W) |
LXCP12-036200 |
3.6 |
2.00 |
3.30~3.90 |
≤3.0 |
7.20
|
LXCP12-042200 |
4.2 |
2.00 |
3.90~4.50 |
≤3.0 |
8.40 |
LXCP12-005100 |
5.0 |
1.00 |
4.70~5.40 |
≤2.0 |
5.0 |
LXCP12-005200 |
5.0 |
2.00 |
4.50~5.50 |
≤3.0 |
10.0 |
LXCP12-006060 |
6.0 |
0.60 |
5.40~6.60 |
≤2.0 |
3.6 |
LXCP12-006100 |
6.0 |
1.00 |
5.40~6.60 |
≤2.0 |
6.0 |
LXCP12-006200 |
6.0 |
2.00 |
5.40~6.60 |
≤3.0 |
12.0 |
LXCP12-007171 |
7.0 |
1.71 |
6.70~7.40 |
≤2.5 |
12.0 |
LXCP12-075160 |
7.5 |
1.60 |
7.20~7.90 |
≤2.5 |
12.0 |
LXCP12-008150 |
8.0 |
1.50 |
7.70~8.40 |
≤2.5 |
12.0 |
LXCP12-009133 |
9.0 |
1.33 |
8.70~9.40 |
≤2.5 |
12.0 |
LXCP12-010120 |
10.0 |
1.20 |
9.70~10.40 |
≤2.0 |
12.0 |
LXCP12-011100 |
11.0 |
1.00 |
10.70~11.40 |
≤2.0 |
11.0 |
LXCP12-012100 |
12.0 |
1.00 |
11.70~12.50 |
≤2.0 |
12.0 |
LXCP12-135088 |
13.5 |
0.88 |
13.00~13.90 |
≤1.5 |
12.0 |
LXCP12-015080 |
15.0 |
0.80 |
14.70~15.40 |
≤1.5 |
12.0 |
Line Regulation |
±2% |
||||
Load Regulation |
±8% |
Ripple and Noise
Under nominal voltage and rated load, the ripple and noise are less than 200mVp-p when measure with Max. Bandwidth of 20MHz and parallel 10uF electrolytic capacitor + 0.1uF ceramic capacitor, connected at testing point.(Output power >6W series)
Under nominal voltage and rated load, the ripple and noise are less than 100mVp-p when measure with Max. Bandwidth of 20MHz and parallel 10uF electrolytic capacitor + 0.1uF ceramic capacitor, connected at testing point.(Output power ≤6W series)
Hold Up Time:
5 ms Min.at 115Vac input and output rated Load.
Efficiency:
70% Min. at 115/230Vac input voltage, rated load.
Protection Function
Short circuit test
The power supply will be automaticly recovering when the short circuit faults removed.
Over current Protection
The power supply will be protected when the output current reaches 110% to 200% of the rated current. Once the overcurrent fault is resolved, the power supply will automatically recover.
Over Voltage Protection
The power supply will shut down when the output voltage reaches its over-voltage condition trigger point. Once the fault condition is resolved, the power supply will automatically recover. The overvoltage protection function is integrated into the PWM IC of the power supply.
Over temperature protection
The power supply will enter a stop mode if the IC's temperature exceeds its trigger point. It will automatically recover when the IC temperature falls below the regulation value.
Environmental Requiremet
Operating Temperature
0℃ to 40℃,Rated load, Normal operation.
Storage Temperature:-20℃ to 80℃
No work
Storage Humidity
5%~95%RH, No condensation.
Atmospheric pressure
54-106KPa, Normal
Altitude
5000m,work temperature descend 1℃ every 300m over 5000m.
(9~200Hz,Acceleration 5m/S2)
Transportation: 5-9Hz, A=3.5mm
Acceleration=5m/S2
Acceleration=15m/S2
Axes, 10 cycles per axis
No permanent damage may occur during testing.
The Power supply can be restore to normal after power off/on.
Dropping Packed
As previously mentioned, the wall mount type requires a 1m distance, while the desktop type requires a 760mm distance.
The horizontal surface should be constructed with hardwood at least 13mm thick, mounted on two layers of plywood, and positioned 19mm to 20mm away from the edge.
Relative Humidity
5%(0℃) ~90%(40℃)RH, 72Hrs, Rated load, Normal operating.
MTBF
The power supply shall have a minimum predicted MTBF(MIL-STD-217F) of 100000 hours at normal operation condition and normal use.
Electromagnetic Compatiblity
Number |
Item |
Specifications |
Class |
Standards |
1 |
CE |
Class B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18; GB4824 |
2 |
RE |
Class B |
/ |
IEC/EN60601-1-2;YY0505 GB4824; EN55011; FCC Part 18 |
3 |
SURGE |
Line to line ± 1KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-5; GB17626.5 |
|
|
Line to GND ± 2KV |
/ |
|
4 |
ESD |
Air discharge ± 15KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-2; GB17626.2 |
|
|
Contact discharge ± 8KV |
A |
|
5 |
EFT/B |
± 2KV (Burst frequency=100KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-4; GB17626.4 |
6 |
DIP |
Dropped to 0%Ut,last 5000ms(250cycle) |
B |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-11 GB17626.11 |
Dropped to 70%Ut,last 500ms(25 cycle) |
A |
|
||
Dropped to 0%Ut,last 20ms(1 cycle) |
A |
|
||
Dropped to0%Ut,last 10ms(0.5 cycle) |
A |
|
||
7 |
RS |
Test frequency:80MHz~2700MHz; Field intensity:10V/m; Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-3 GB17626.3 |
8 |
CS |
Test frequency:0.15MHz~80MHz; Field intensity:6Vrms; Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-6 GB17626.6 |
9 |
THD |
Class A(in system) |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-2; GB17625.1 |
10 |
Voltage Fluctuation and Flicker |
Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (Dmax) under4% |
/ |
IEC/EN60601-1-2 YY0505 IEC/EN61000-3-3 GB17625.2 |
11 |
Power Frequency Magnetic Field |
30A/m |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-8; GB17626.8 |
Safety: accord with
Item |
Country |
Standard |
|
□ |
UL |
USA |
UL60950-1/UL60601-1 |
■ |
CE |
Europe |
EN60950-1/EN60601-1 |
■ |
CB |
Worldwide |
IEC60601-1 |
■ |
TUV |
Germany |
IEC60601-1 |
■ |
NRTL |
CUSA |
IEC60601-1/UL60601-1 |
□ |
GS |
Germany |
EN60601-1 |
□ |
BS |
England |
EN60601-1 |
□ |
SAA |
Australia |
AS/NZS6-1 |
Mechanical Requirement
Optional input plug:
Cable
label size:24.5mm*17.5mm
Company Profile
According to the needs of different customers to customize the development of new products, in addition to the conventional research and development equipment configuration, we are very few in the industry with EMC radio wave room, safety laboratory, RoHS, reliability laboratory and other facilities. Our qualification certification is complete, and has a professional after-sales service team, to provide 7*24 hours online response service, in case of special circumstances can be on call. High quality products are derived from advanced and reliable design, high standard requirements of components and rigorous production process management; Based on the continuous tracking of domestic and foreign advanced technology dynamics, independent R & D product verification department as a guarantee, through the team's review (DFX, DFMEA, etc.) to get high-quality product design.