EN 60601-1 Medical Grade Power Supply 9V 12V 15V 19V 24V DC Single Switching Adapter Manufacturer LXCP150
Input Characteristics
Input Voltage:
Nominal Voltage: 100~240Vac
Variation Range: 90~264Vac
Input Frequency:
Nominal Frequency: 50/60Hz.
Variation Frequency: 47~63Hz
Input Current:
0.3Amps max at any input voltage and rated DC output and rated load.
Inrush Current:
50Amps Max. Cold start at 264Vac input, with rated load and 25℃ ambient.
AC Leakage Current:
Normal 0.1mA Max.at 264Vac input.
Single fault 0.2mA Max.at 264Vac input.
Output Characteristics
Model name |
Output Voltage(V) |
Rated Load(A) |
Output Range(V) |
Output power(W) |
LXCP6-036 |
3.6 |
1.20 |
3.10~4.00 |
6.0 |
LXCP6-042 |
4.2 |
1.20 |
3.70~4.60 |
6.0 |
LXCP6-050 |
5.0 |
1.20 |
4.40~5.60 |
6.0 |
LXCP6-060 |
6.0 |
1.00 |
5.40~6.60 |
6.0 |
LXCP6-075 |
7.5 |
0.80 |
7.00~8.10 |
6.0 |
LXCP6-084 |
8.4 |
0.71 |
7.90~8.90 |
6.0 |
LXCP6-090 |
9.0 |
0.66 |
8.50~9.70 |
6.0 |
LXCP6-100 |
10.0 |
0.60 |
9.50~10.70 |
6.0 |
LXCP6-120 |
12.0 |
0.50 |
11.30~12.70 |
6.0 |
LXCP6-126 |
12.6 |
0.50 |
12.00~13.30 |
6.3 |
Line Regulation |
±3% |
|||
Load Regulation |
±5% |
Ripple and Noise
Test conditions: Under nominal voltage and rated load, the ripple and noise are less than 350mVp-p when measure with Max. Bandwidth of 20MHz and parallel 10uF/0.1uF, connected at testing point.
Turn on delay time
2Second Max.at 220Vac input and output rated load.
Rise time
30mS Max.at 115Vac input and output rated load.
Hold up time
5mS Min.at 115Vac input and output rated Load.
Efficiency:
70% Min, at 115/230Vac input voltage, rated load,Test USB end of power supply
Protection Function
Short circuit test
The power supply will be automaticity recovery when short circuit faults remove.
Over current Protection
When the output current exceeds 110% to 200% of the rated current, the power supply will activate protection measures. Once the overcurrent fault is resolved, the power supply will automatically resume normal operation.
Over Voltage Protection
When the output voltage reaches 105%~125% of the rated voltage, the power supply will be protected and can resume normal work after the fault removed.
Over temperature protection
5mS Min.at 115Vac input and output rated Load.
The power supply will enter a stop mode if the IC’s temperature exceeds its trigger point. When the IC temperature falls below the regulation value, the power supply will automatically recover.
Environmental Requiremet
Operating Temperature
0℃ to 40℃,Rated load, Normal operation.
Storage Temperature:-20℃ to 80℃
No work
Storage Humidity: 10%~90%
No condensation
Atmospheric pressure
70-106KPa, Normal.
Altitude
5000m,Work temp descend 1℃ every 300m over 5000m.
(9~200Hz,Acceleration 5m/S2)
Transportation: 5-9Hz, A=3.5mm
Acceleration=5m/S2
Acceleration=15m/S2
Axes, 10 cycles per axis
No permanent damage may occur during testing.
The Power supply can be restore to normal after power off/on.
Dropping Packed
The wall mount type requires a 1m distance, while the desktop type requires a 760mm distance as described above.
The horizontal surface should be made of hardwood at least 13mm thick, mounted on two layers of plywood, and positioned 19mm to 20mm away from the edge.
Relative Humidity
5%(0℃) ~90%(40℃)RH, 72Hrs, Rated load, Normal operating.
MTBF
The power supply is expected to have a minimum MTBF (MIL-STD-217F) of 100,000 hours under normal operating conditions and typical usage.
Electromagnetic Compatiblity
Numbe |
Item |
Specifications |
Class |
Standards |
1 |
(CE) |
CLASS B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18 |
2 |
(RE) |
CLASS B |
/ |
IEC/EN60601-1-2; YY0505 GB4824; EN55011; FCC Part 18 |
3 |
(SURGE) |
Line to line±1KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-5; GB17626.5 |
|
|
Line to GND±2KV |
A |
|
4 |
(ESD) |
Air discharge±15KV |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-2; GB17626.2 |
|
|
Contact discharge±8KV |
A |
|
5 |
(EFT/B) |
±2KV (BURST FREQUENCY=100KHZ) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-4; GB17626.4 |
6 |
(DIP) |
Dropped to 0%Ut,last 5000ms(250cycle) |
B |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-11; GB17626.11 |
Dropped to 30%Ut,last 500ms(25 cycle) |
B |
|
||
Dropped to 0%Ut,last 20ms(1 cycle) |
B |
|
||
Dropped to0%Ut,last 10ms(0.5 cycle) |
A |
|
||
7 |
(RS) |
Test frequency:80MHz~2700MHz; Field intensity:10V/m;80%AM(1KHz) Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-3; GB17626.3 |
8 |
(CS) |
Test frequency:0.15MHz~80MHz; Field intensity:6Vrms; Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-6; GB17626.6 |
9 |
(THD) |
CLASSA(in system) |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-2; GB17625.1 |
10 |
Voltage Fluctuation and Flicker |
Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4% |
/ |
IEC/EN60601-1-2; YY0505 IEC/EN61000-3-3; GB17625.2 |
11 |
Power Frequency Magnetic Field |
30A/m |
A |
IEC/EN60601-1-2; YY0505 IEC/EN61000-4-8; GB17626.8 |
Safety: accord with
ITEM |
COUNTRY |
STANDARD |
|
□ |
UL |
USA |
UL60950-1/UL60601-1 |
■ |
CE |
EUROPE |
EN60950-1/EN60601-1 |
■ |
CB |
WORLDWIDE |
IEC60601-1 |
□ |
TUV |
GERMANY |
IEC60601-1 |
■ |
NRTL |
CUSA |
IEC60601-1/UL60601-1 |
□ |
GS |
GERMANY |
EN60601-1 |
□ |
BS |
ENGLAND |
EN60601-1 |
□ |
SAA |
AUSTRALIA |
AS/NZS6-1 |