EN 60601-1 Medical Grade Power Supply 9V 12V 15V 19V 24V DC Single Switching Adapter Manufacturer LXCP150

·Medical Certifications: Medical adapters are certified to meet rigorous safety standards, such as IEC 60601-1, which ensures they are safe for use in medical environments, especially in patient-connected devices.
·Low Leakage Current: They are designed with low leakage current to minimize the risk of electrical shocks, which is critical in environments where patient safety is paramount.

    Input Characteristics

    Input Voltage:

    Nominal Voltage: 100~240Vac

    Variation Range: 90~264Vac

    Input Frequency:

    Nominal Frequency: 50/60Hz.

    Variation Frequency: 47~63Hz

    Input Current:

    0.3Amps max at any input voltage and rated DC output and rated load.

    Inrush Current:

    50Amps Max. Cold start at 264Vac input, with rated load and 25℃ ambient.

    AC Leakage Current:

    Normal 0.1mA Max.at 264Vac input.

    Single fault 0.2mA Max.at 264Vac input.

    Output Characteristics

    Model name

    Output Voltage(V)

    Rated Load(A)

    Output Range(V)

    Output power(W)

    LXCP6-036

    3.6

    1.20

    3.10~4.00

    6.0

    LXCP6-042

    4.2

    1.20

    3.70~4.60

    6.0

    LXCP6-050

    5.0

    1.20

    4.40~5.60

    6.0

    LXCP6-060

    6.0

    1.00

    5.40~6.60

    6.0

    LXCP6-075

    7.5

    0.80

    7.00~8.10

    6.0

    LXCP6-084

    8.4

    0.71

    7.90~8.90

    6.0

    LXCP6-090

    9.0

    0.66

    8.50~9.70

    6.0

    LXCP6-100

    10.0

    0.60

    9.50~10.70

    6.0

    LXCP6-120

    12.0

    0.50

    11.30~12.70

    6.0

    LXCP6-126

    12.6

    0.50

    12.00~13.30

    6.3

    Line Regulation

    ±3%

    Load Regulation

    ±5%

    Ripple and Noise

    Test conditions: Under nominal voltage and rated load, the ripple and noise are less than 350mVp-p when measure with Max. Bandwidth of 20MHz and parallel 10uF/0.1uF, connected at testing point. 

    Turn on delay time

    2Second Max.at 220Vac input and output rated load.

    Rise time

    30mS Max.at 115Vac input and output rated load.

    Hold up time

    5mS Min.at 115Vac input and output rated Load.

    Efficiency:

    70% Min, at 115/230Vac input voltage, rated load,Test USB end of power supply

    Protection Function

    Short circuit test

    The power supply will be automaticity recovery when short circuit faults remove.

    Over current Protection

    When the output current exceeds 110% to 200% of the rated current, the power supply will activate protection measures. Once the overcurrent fault is resolved, the power supply will automatically resume normal operation.

    Over Voltage Protection

    When the output voltage reaches 105%~125% of the rated voltage, the power supply will be protected and can resume normal work after the fault removed.

    Over temperature protection

    5mS Min.at 115Vac input and output rated Load.

    The power supply will enter a stop mode if the IC’s temperature exceeds its trigger point. When the IC temperature falls below the regulation value, the power supply will automatically recover.

    Environmental Requiremet

    Operating Temperature

    0℃ to 40℃,Rated load, Normal operation.

    Storage Temperature:-20℃ to 80℃

    No work

    Storage Humidity: 10%~90%

    No condensation

    Atmospheric pressure

    70-106KPa, Normal.

    Altitude

    5000m,Work temp descend 1℃ every 300m over 5000m.

    (9~200Hz,Acceleration 5m/S2)

    Transportation: 5-9Hz, A=3.5mm

    Acceleration=5m/S2

    Acceleration=15m/S2

    Axes, 10 cycles per axis

    No permanent damage may occur during testing.

    The Power supply can be restore to normal after power off/on.

    Dropping Packed

    The wall mount type requires a 1m distance, while the desktop type requires a 760mm distance as described above.

    The horizontal surface should be made of hardwood at least 13mm thick, mounted on two layers of plywood, and positioned 19mm to 20mm away from the edge.

    Relative Humidity

    5%(0℃) ~90%(40℃)RH, 72Hrs, Rated load, Normal operating.

    MTBF

    The power supply is expected to have a minimum MTBF (MIL-STD-217F) of 100,000 hours under normal operating conditions and typical usage.

    Electromagnetic Compatiblity

    Numbe

    Item

    Specifications

    Class

    Standards

    1

    (CE)

    CLASS B

    /

    IEC/EN60601-1-2; YY0505

    GB4824; EN55011;

    FCC Part 18

    2

    (RE)

    CLASS B

    /

    IEC/EN60601-1-2; YY0505

    GB4824; EN55011;

    FCC Part 18

    3

    (SURGE)

    Line to line±1KV

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-5; GB17626.5

     

     

    Line to GND±2KV

    A

     

    4

    (ESD)

    Air discharge±15KV

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-2; GB17626.2

     

     

    Contact discharge±8KV

    A

     

    5

    (EFT/B)

    ±2KV (BURST FREQUENCY=100KHZ)

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-4; GB17626.4

    6

    (DIP)

    Dropped to 0%Ut,last 5000ms(250cycle)

    B

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-11; GB17626.11

    Dropped to 30%Ut,last 500ms(25 cycle)

    B

     

    Dropped to 0%Ut,last 20ms(1 cycle)

    B

     

    Dropped to0%Ut,last 10ms(0.5 cycle)

    A

     

    7

    (RS)

    Test frequency:80MHz~2700MHz;

    Field intensity:10V/m;80%AM(1KHz)

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-3; GB17626.3

    8

    (CS)

    Test frequency:0.15MHz~80MHz;

    Field intensity:6Vrms;

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-4-6; GB17626.6

    9

    (THD)

    CLASSA(in system)

    /

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-3-2; GB17625.1

    10

    Voltage Fluctuation and Flicker

    Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4%

    /

    IEC/EN60601-1-2;

    YY0505

    IEC/EN61000-3-3; GB17625.2

    11

    Power Frequency Magnetic Field

    30A/m

    A

    IEC/EN60601-1-2; YY0505

    IEC/EN61000-4-8; GB17626.8

    Performance criterion A: Normal performance within limits in this specification. 
    Performance criterion B: Temporary loss of some functions or degradation of performance. The recovering performance of which without operator intervention.
    Performance Criterion C: Temporary loss of function or performance degradation that requires operator intervention for recovery.
    Performance Criterion D: Loss of function or performance degradation that is not recoverable due to hardware or software damage, or data loss.

    Safety: accord with

    The power supply is intended for indoor use and is designed to comply with the safety standards of IEC 60601-1 and EN 60601-1, as well as other relevant safety regulations.
     

    ITEM

    COUNTRY

    STANDARD

    UL

    USA

    UL60950-1/UL60601-1

    CE

    EUROPE

    EN60950-1/EN60601-1

    CB

    WORLDWIDE

    IEC60601-1

    TUV

    GERMANY

    IEC60601-1

    NRTL

    CUSA

    IEC60601-1/UL60601-1

    GS

    GERMANY

    EN60601-1

    BS

    ENGLAND

    EN60601-1

    SAA

    AUSTRALIA

    AS/NZS6-1

    The primary to secondary: 4000Vac 10mA for 60 seconds

    Mechanical Requirement

    EU ping the power supply size: L 78.0 x W 35.6 x H 24.1mm; US ping the power supply size: L 40.5 x W 34.7 x H 24.1mm;
    wiling3ehs
    CN ping the power supply size: L 40.5 x W 34.7 x H 24.1mm;
    Enterprise Honor014ik

    Cable

    U@SH{S4JHKNC`BZPFY0XJC90n3

    optional DC plug 

    optional DC plug  (1)skdoptional DC plug  (2)j9doptional DC plug  (3)k9yoptional DC plug  (4)1kd
    Note: Details are selected from the LXC_ Wire Library drawing XLS file

    Label

    Label (1)x0hLabel (2)8ihLabel (3)mmd

    FAq

    1. What is the scale of the company and factory? 
    Our company has professional consultants and engineers, 25 after-sales service teams and more than 250 factory production management personnel.  
    2. What type of medical power supply does Longxc mainly produce? 
    From 1-600W power supply. Our medical power supply cover monitoring, anesthesia, respiratory, electrocardiogram machine, infusion pump, injection pump, B-ultrasound, imaging, biochemistry, electronic blood pressure monitor, beauty, rehabilitation physiotherapy and other medical equipment
    3. What are the advantages of company? 
    A. Our professional design and engineering team with more than 15 years of experience can provide professional advice and solutions to customers and efficiently solve any problems that arise.                         
    B: We are a professional medical power supply factory with 15 years of experience, providing you with the most cost-effective products and saving your money.  
    C: Own factories provide more flexibility and can prioritize our direct customer orders to save customers valuable time.
    D: Senior medical power supply company, with excellent reputation in the medical power supply industry, regards integrity as life, and ensures 100% customer capital and information security.     
    4.What are the most representative customers of the company's brands or projects? 
    Our clients we have served and are cooperating with include: mindray, cardinalhealth, sino,Creative, Fresenius  etc. 
    5. How long is the company's product warranty? 
    The regular warranty period of our products is 36 months, and some products can be guaranteed for  more than 5years.
    6. What are the customer groups of the company? 
    Our customers are mainly domestic and foreign.
    7. What are the company's payment terms? 
    Our payment terms are 30% deposit and the balance shall be paid before shipment.d catering chain enterprises, etc.
    8. Can you provide samples? 
    Our company can provide samples for customers to confirm the quality.
    9. Can I visit the factory? 
    Our factory is located in Shenzhen, Guangdong Province. We welcome friends from all over the world to visit our factory and discuss cooperation at any time.