50W DC 12V4A Medical fental monitor Power Adapter 50W Desktop Type C6 C8 C14 Inlet LXCP52
INPUT VOLTAGE & FREQUENCY
Input Current:
The maximum input current is 1.5A at nominal AC input voltage and full load.
Inrush Current:
The inrush current will not exceed 50A at 264Vac input for a cold start at 25℃.
STATIC LOSS
The static loss is less than 0.5W at normal input voltage.
The efficiency shall be higher than 80% while measuring at 115VAC&230VAC input voltage, full load.
Output Characteristics
Item |
Model number |
OUTPUT CHARACTERISTICS |
||||
|
LXCP52 series |
LXCP52(II) series |
Rated output voltage(V) |
Output voltage range(V) |
Rated output current(A) |
Rated output power(W) |
Model number |
LXCP52-005 |
LXCP52(II)-005 |
5.0 |
4.50~5.40 |
4.00 |
20.0 |
|
LXCP52-009 |
LXCP52(II)-009 |
9.0 |
8.50~9.40 |
4.00 |
36.0 |
|
LXCP52-012 |
LXCP52(II)-012 |
12.0 |
11.60~12.50 |
4.00 |
48.0 |
|
LXCP52-014 |
LXCP52(II)-014 |
14.0 |
13.60~14.50 |
3.30 |
46.2 |
|
LXCP52-015 |
LXCP52(II)-015 |
15.0 |
14.50~15.50 |
3.30 |
49.5 |
|
LXCP52-168 |
LXCP52(II)-168 |
16.8 |
16.50~17.00 |
2.63 |
44.2 |
|
LXCP52-018 |
LXCP52(II)-018 |
18.0 |
17.80~18.45 |
2.63 |
47.4 |
|
LXCP52-019 |
LXCP52(II)-019 |
19.0 |
18.62~19.38 |
2.63 |
50.0 |
|
LXCP52-020 |
LXCP52(II)-020 |
20.0 |
19.60~20.40 |
2.20 |
44.0 |
|
LXCP52-024 |
LXCP52(II)-024 |
24.0 |
23.50~24.50 |
2.20 |
52.8 |
load regulation |
±2% |
|||||
voltage regulation |
±2% |
|||||
overshoot |
≤5% |
|||||
Hold up time |
≥16 mS |
|||||
Efficiency |
≥80% |
|||||
Over current protection(A) |
110-200%Iout |
|||||
Over voltage protection(V) |
105-140%Vout |
|||||
Ripple and noise |
≤200mV (measured by 20MHz bandwidth oscilloscope and terminated each output with a 0.1uF ceramic capacitor and a10uF aluminum electrolytic capacitor) |
|||||
Over temperature protection |
The power supply shall go into stop mode when the IC's temperature is over its trigger point and when IC temperature is lower than regulation value, the power supply shall be self-recovering. |
|||||
Short circuit protection |
The power supply shall withstand continuous current of short circuit on output no any damage. When the fault condition is removed the power supply shall be self-recovering. |
OPERATING CONDITION
The power supply shall be compliant with each item in this specification for the following environmental conditions.
Ambient Temperature |
-10℃~40℃ |
Relative Humidity |
20%~90% |
STORAGE CONDITION
Ambient Temperature |
-20℃~80℃ |
Relative Humidity |
10%~90% |
Rise time
30mS Max.at 115Vac input and output rated load.
Hold up time
5mS Min.at 115Vac input and output rated Load.
MTBF
The power supply shall have a minimum predicted MTBF(MIL-STD-217F) of 100000 hours at normal operation condition and normal use.
Protection Function
Short circuit test
The power supply will be automaticity recovery when short circuit faults remove.
Over current Protection
When the output current reaches 110%~200% of the rated current, the power supply will be protected. After the overcurrent fault is removed, the power supply will be automatically recovery.
Over Voltage Protection
When the output voltage reaches 105%~125% of the rated voltage, the power supply will be protected and can resume normal work after the fault removed.
Over temperature protection
5mS Min.at 115Vac input and output rated Load.
The power supply shall go into stop mode when the IC's temperature is over its trigger point and when IC temperature is lower than regulation value, the power supply shall be self-recovering.
Electromagnetic Compatiblity
Number |
Item |
Specifications |
Class |
Standards |
1 |
CE |
Class B |
/ |
IEC/EN60601-1-2 YY0505 GB4824 EN55011 FCC Part 18 GB4824 |
2 |
RE |
Class B |
/ |
IEC/EN60601-1-2 YY0505 GB4824 EN55011 FCC Part 18 |
3 |
SURGE |
Line to line ± 1KV |
A |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-5 GB17626.5 |
Line to GND ± 2KV |
A |
|
||
4 |
ESD |
Air discharge ± 15KV |
A |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-2 GB17626.2 |
Contact discharge ± 8KV |
A |
|
||
5 |
EFT/B |
±2KV (Burst frequency = 100KHz) |
A |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-4 GB17626.4
|
6 |
DIP |
Dropped to 0%Ut,last 5000ms(250cycle) |
B |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-11 GB17626.11 |
Dropped to 30%Ut,last 500ms(25 cycle) |
B |
|
||
Dropped to 0%Ut,last 20ms(1 cycle) |
B |
|
||
Dropped to 0%Ut,last 10ms(0.5 cycle) |
A |
|
||
7 |
RS |
Test frequency:80MHz~2700MHz; Field intensity:10V/m; Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-3 GB17626.3 |
8 |
CS |
Test frequency:0.15MHz~80MHz; Field intensity:6V; Amplitude modulation:80%AM(1KHz) |
A |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-6 GB17626.6 |
9 |
THD |
Class A(in system) |
/ |
IEC/EN60601-1-2 YY0505 IEC/EN61000-3-2 GB17625.1 |
10 |
Voltage Fluctuation and Flicker |
Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4% |
/ |
IEC/EN60601-1-2 YY0505 IEC/EN61000-3-3 GB17625.2 |
11 |
Power Frequency Magnetic Field |
30A/m |
A |
IEC/EN60601-1-2 YY0505 IEC/EN61000-4-8 GB17626.8 |
Safety: accord with
ITEM |
COUNTRY |
STANDARD |
|
□ |
UL |
USA |
UL60950-1/UL60601-1 |
■ |
CE |
EUROPE |
EN60950-1/EN60601-1 |
■ |
CB |
WORLDWIDE |
IEC 60601-1 |
■ |
TUV |
GERMANY |
IEC60601-1 |
□ |
NRTL |
CUSA |
IEC60601-1/UL6001-1 |
□ |
GS |
GRMANY |
EN60601-1 |
□ |
BS |
ENGLAND |
EN60601-1 |
□ |
SAA |
AUSTRALIA |
AS/NZS6-1 |