Leave Your Message
Products Categories
Featured Products

50W DC 12V4A Medical fental monitor Power Adapter 50W Desktop Type C6 C8 C14 Inlet LXCP52

This document introduces the electrical, mechanical and environmental specifications of a switching power supply.

    INPUT VOLTAGE & FREQUENCY

    Input Current:

    The maximum input current is 1.5A at nominal AC input voltage and full load.

    Inrush Current:

    The inrush current will not exceed 50A at 264Vac input for a cold start at 25℃.

    STATIC LOSS

    The static loss is less than 0.5W at normal input voltage.

    The efficiency shall be higher than 80% while measuring at 115VAC&230VAC input voltage, full load.

    Output Characteristics

    Item

    Model number

    OUTPUT CHARACTERISTICS

     

    LXCP52

    series

    LXCP52(II)

    series

    Rated output voltage(V)

    Output voltage range(V)

    Rated output current(A)

    Rated output power(W)

    Model number

    LXCP52-005

    LXCP52(II)-005

    5.0

     4.50~5.40

    4.00

    20.0

     

    LXCP52-009

    LXCP52(II)-009

    9.0

     8.50~9.40

    4.00

    36.0

     

    LXCP52-012

    LXCP52(II)-012

    12.0

    11.60~12.50

    4.00

    48.0

     

    LXCP52-014

    LXCP52(II)-014

    14.0

    13.60~14.50

    3.30

    46.2

     

    LXCP52-015

    LXCP52(II)-015

    15.0

    14.50~15.50

    3.30

    49.5

     

    LXCP52-168

    LXCP52(II)-168

    16.8

    16.50~17.00

    2.63

    44.2

     

    LXCP52-018

    LXCP52(II)-018

    18.0

    17.80~18.45

    2.63

    47.4

     

    LXCP52-019

    LXCP52(II)-019

    19.0

    18.62~19.38

    2.63

    50.0

     

    LXCP52-020

    LXCP52(II)-020

    20.0

    19.60~20.40

    2.20

    44.0

     

    LXCP52-024

    LXCP52(II)-024

    24.0

    23.50~24.50

    2.20

    52.8

    load regulation

    ±2%

    voltage regulation

    ±2%

    overshoot

    ≤5%

    Hold up time

    ≥16 mS

    Efficiency

    ≥80%

    Over current protection(A)

    110-200%Iout

    Over voltage protection(V)

    105-140%Vout

    Ripple and noise

    ≤200mV (measured by 20MHz bandwidth oscilloscope and terminated each output with a 0.1uF ceramic capacitor and a10uF aluminum electrolytic capacitor)

    Over temperature protection

    The power supply shall go into stop mode when the IC's temperature is over its trigger point and when IC temperature is lower than regulation value, the power supply shall be self-recovering.

    Short circuit protection

    The power supply shall withstand continuous current of short circuit on output no any damage. When the fault condition is removed the power supply shall be self-recovering.

    OPERATING CONDITION

    The power supply shall be compliant with each item in this specification for the following environmental conditions.

    Ambient Temperature

    -10℃~40℃

    Relative Humidity

    20%~90%

    STORAGE CONDITION

    Ambient Temperature

    -20℃~80℃

    Relative Humidity

    10%~90%

    Rise time

    30mS Max.at 115Vac input and output rated load.

    Hold up time

    5mS Min.at 115Vac input and output rated Load.

    MTBF

    The power supply shall have a minimum predicted MTBF(MIL-STD-217F) of 100000 hours at normal operation condition and normal use.

    Protection Function

    Short circuit test

    The power supply will be automaticity recovery when short circuit faults remove.

    Over current Protection

    When the output current reaches 110%~200% of the rated current, the power supply will be protected. After the overcurrent fault is removed, the power supply will be automatically recovery.

    Over Voltage Protection

    When the output voltage reaches 105%~125% of the rated voltage, the power supply will be protected and can resume normal work after the fault removed.

    Over temperature protection

    5mS Min.at 115Vac input and output rated Load.

    The power supply shall go into stop mode when the IC's temperature is over its trigger point and when IC temperature is lower than regulation value, the power supply shall be self-recovering.

    Electromagnetic Compatiblity

    Number

    Item

    Specifications

    Class

    Standards

    1

    CE

    Class B

    /

    IEC/EN60601-1-2

    YY0505

    GB4824

    EN55011

    FCC Part 18

    GB4824

    2

    RE

    Class B

    /

    IEC/EN60601-1-2

    YY0505

    GB4824

    EN55011

    FCC Part 18

    3

    SURGE

    Line to line ± 1KV

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-5

    GB17626.5

    Line to GND ± 2KV

    A

     

    4

    ESD

    Air discharge ± 15KV

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-2

    GB17626.2

    Contact discharge ± 8KV

    A

     

    5

    EFT/B

    ±2KV (Burst frequency = 100KHz)

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-4

    GB17626.4

     

    6

    DIP

    Dropped to 0%Ut,last 5000ms(250cycle)

    B

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-11

    GB17626.11

    Dropped to 30%Ut,last 500ms(25 cycle)

    B

     

    Dropped to 0%Ut,last 20ms(1 cycle)

    B

     

    Dropped to 0%Ut,last 10ms(0.5 cycle)

    A

     

    7

    RS

    Test frequency:80MHz~2700MHz;

    Field intensity:10V/m;

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-3

    GB17626.3

    8

    CS

    Test frequency:0.15MHz~80MHz;

    Field intensity:6V;

    Amplitude modulation:80%AM(1KHz)

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-6

    GB17626.6

    9

    THD

    Class A(in system)

    /

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-3-2

    GB17625.1

    10

    Voltage

    Fluctuation  and Flicker

    Pst≤1.0;Plt≤0.65;Relative steady-state voltage variation dc under3.3%;Maximum relative voltage variation (dmax) under4%

    /

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-3-3

    GB17625.2

    11

    Power Frequency Magnetic Field

    30A/m

    A

    IEC/EN60601-1-2

    YY0505

    IEC/EN61000-4-8

    GB17626.8

    Performance criterion A: Normal performance within limits in this specification. 
    Performance criterion B: Temporary loss of some functions or degradation of performance. The recovering performance of which without operator intervention.
    Performance criterion C: Temporary loss of function or degradation of performance .The recovering performance of which requires operator intervention.
    Performance criterion D: Loss of function or degradation of performance which is not recoverable .Owing to damage of hardware software, or loss of data.

    Safety: accord with

    The power supply is designed for indoor use to meet IEC 60601-1 and EN 60601-1standards and for the following safety regulations.
     

    ITEM

    COUNTRY

    STANDARD

    UL

    USA

    UL60950-1/UL60601-1

    CE

    EUROPE

    EN60950-1/EN60601-1

    CB

    WORLDWIDE

    IEC 60601-1

    TUV

    GERMANY

    IEC60601-1

    NRTL

    CUSA

    IEC60601-1/UL6001-1

    GS

    GRMANY

    EN60601-1

    BS

    ENGLAND

    EN60601-1

    SAA

    AUSTRALIA

    AS/NZS6-1

    DIELECTRIC WITHSTAND VOLTAGE
    Primary to Secondary: 4KVac for 60 seconds, 5mA(max)
    LEAKAGE CURRENT
    According to the class II with Leakage current less than 0.1mA

    Mechanical Requirement

    L*W*H   112.0*47.5*33.5
    asd 4or3

    optional DC plug 

    optional DC plug  (1)skdoptional DC plug  (2)j9doptional DC plug  (3)k9yoptional DC plug  (4)1kd
    Optional input plug: The AC input line shall be required for VDE or UL certification.

    NAMEPLATE 

    0wnp图片 11qf8