0102030405
150W Desktop dc adapter 15V 10A 150W power supply 15v 10 amp power supply for medical device LXCP120
parameter
Model number |
Model number |
Output current range |
LXCP120-012bbbb |
LXCP120(II)-012bbbb |
bbbb="0100"-"1000",1.0A-10.0 A step by 0.01A |
LXCP120-015bbbb |
LXCP120(II)-015bbbb |
bbbb="0100"-"1000",1.0A-10.0 A step by 0.01A |
LXCP120-018bbbb |
LXCP120(II)-018bbbb |
bbbb="0100"-"0725",1.0A-7.25 A step by 0.01A |
LXCP120-019bbbb |
LXCP120(II)-019bbbb |
bbbb="0100"-"0625",1.0A-6.75A step by 0.01A |
LXCP120-024bbbb |
LXCP120(II)-024bbbb |
bbbb="0100"-"0625",1.0A-6.25A step by 0.01A |
LXCP120-028bbbb |
LXCP120(II)-028bbbb |
bbbb="0100"-"0535",1.0A-5.35A step by 0.01A |
LXCP120-036bbbb |
LXCP120(II)-036bbbb |
bbbb="0100"-"0415",1.0A-4.15A step by 0.01A |
LXCP120-048bbbb |
LXCP120(II)-048bbbb |
bbbb="0100"-"0315",1.0A-3.15A step by 0.01A |
This document introduces the electrical, mechanical and environmental specifications of a switching power supply.
2 MODEL NAME RULE
LXCP120-aaabbbb
LXCP120(II)-aaabbbb:
-LXCP= Longxc medical adapter
-120=Name of the series
-(II)= Class II equipment
- aaa=“012”-“048” stands for the output condition. “012” means 12.0V, “015” means 15.0V, “018” means 18.0V, “019”
means 19.0V, “024” means 24.0V, “028” means 28.0V, “036” means 36.0V,“048” means 48.0V.
- bbbb=4 digits with range of "0100"-"1000" step by 1, which stands for output current. "0100" means 1.0A, "1000"
means 10.0A. Details refer to the below table:
LXCP120-aaabbbb: For Class I Product, the Earth leakage current less than 300u A at 264Vac,60Hz.
LXCP120(II)-aaabbbb: For Class II Product , the Earth leakage current less than 100u A at 264Vac,60Hz.
3 INPUT CHARACTERISTICS
3.1 INPUT VOLTAGE & FREQUENCY
Minimum Nominal Maximum
Input Voltage 90Vac 100-240Vac 264Vac
Input Frequency 47Hz 50Hz/60Hz 63Hz
3.2 INPUT CURRENT
The maximum input current is 2.5A at nominal AC input voltage and full load.
3.3 INRUSH CURRENT
The inrush current will not exceed 50A at 264Vac input for a cold start at 25℃.
3.4 STATIC LOSS
The static loss is less than 0.5W at normal input voltage.
3.5 EFFICIENCY
The efficiency shall be higher than 90% while measuring at 115VAC&230VAC input voltage, full load.
L*W*H 172.0*72.0*42.0
11 DC PLUG FIGURE(Detail information selection from file of LXC_Dcline_Library.xls)
11 DC PLUG FIGURE(Detail information selection from file of LXC_Dcline_Library.xls)